Skip to main content

Table 5 Summary of mean differences between intraoperative and image-based measurements. P-value is relative to the mean delta being different from zero

From: CT validation of intraoperative imageless navigation (Naviswiss) for component positioning accuracy in primary total hip arthroplasty in supine patient position: a prospective observational cohort study in a single-surgeon practice

Sign

Mean Delta

SE

SD

95%LCI

95%UCI

P-value

Lower LOA

Upper LOA

Inclination

-1.9

0.77

4.5

-3.4

0.4

0.015

-10.7

6.9

Version

-3.6

0.63

3.7

-4.8

-2.4

<0.001

-10.8

3.6

Offset

1.5

0.85

5.0

-0.1

3.2

0.069

-8.2

11.2

Leg Length

2.1

0.76

4.4

0.5

3.6

0.009

-6.6

10.8

Absolute

 Inclination

3.3

0.61

3.6

2.1

4.5

   

 Version

4.3

0.51

3.0

3.3

5.2

   

 Offset

3.9

0.56

3.3

2.8

5

   

 Leg Length

3.7

0.55

3.2

2.7

4.8

   
  1. LOA limits of agreement, LCI: lower confidence interval, UCI upper confidence interval, SE standard error of the estimate, SD standard deviation