Skip to main content

Table 3 Linear fit of average of measurements to delta of measurements

From: Imageless navigation system (Naviswiss) provides accurate component position in total hip arthroplasty with lateral decubitus position for end-stage hip osteoarthritis: a prospective cohort study with CT-validation

 

Coefficient

SE

95% LCI

95% UCI

P-Value

Adjusted R2

Inclination

-1.51

0.1

-1.70

-1.31

 <0.001

0.70

Version

-0.2

0.23

-0.65

0.24

0.368

0.00

Offset

-0.24

0.15

-0.53

0.05

0.102

0.04

LLD

-0.51

0.09

-0.69

-0.32

 <0.001

0.41