Open reduction and internal fixation with cables for the variant AGT Periprosthetic fracture: a case report and literature review

A Correction to this article was published on 05 May 2020

This article has been updated

Abstract

Background

Periprosthetic femoral fracture is identified as the third most frequent reason for revision total hip arthroplasty (THA). Treatment of periprosthetic fractures of the femur after THA remains a surgical challenge. In this report, we presented 2 patients with periprosthetic proximal femur fracture variant (a fracture of the greater trochanter with lateral cortical extension) and femoral stem destabilization.

Cases presentation

Two patients presented with chief complaints of pain in hip, restricted hip movements and gait changes. On the basis of clinicoradiological findings, the patients were diagnosed as pseudo AGT periprosthetic fracture, since the stem was loosened. They underwent open reduction and internal fixation (ORIF) with cables. After 2 years of follow-up, the 2 patients had favorable clinical outcomes after operation. Both lower limbs of the 2 patients were of equal length. The Harris score of the two hips was 96 and 94, respectively.

Conclusion

CT scan worked better than X-ray examination in the diagnosis of prosthetic looseness with this type of fracture. Compared to longer-stem revision, ORIF with cables could also achieve good result with these fractures.

Background

Periprosthetic femoral fracture (PPFF) is increasingly becoming a common complication of total hip arthroplasty (THA) and identified to be the third most frequent reason for revision THA [1]. With primary THA, the rate of intra-operative PPFF was reportedly 1.7% and a 20-year follow-up showed that the long-term rate was 3.5% [2].

Periprosthetic fractures are difficult to manage and may lead to poor outcomes. Post-THA treatment of PPFF remains a surgical challenge [3,4,5].

Presented in this report, were 2 patients suffering from periprosthetic proximal femur fracture variant (a fracture of the greater trochanter with lateral cortical extension), and femoral stem destabilization. The 2 patients had favorable clinical outcomes after open reduction and internal fixation (ORIF) with cables. Consents were obtained from the patients after they had been informed the fact their pictures might be submitted for publication.

Case series

Case 1

A 69-year-old man tripped and fell over. Subsequently, persistent pain developed in the right hip. Bilateral radiography of hips 2 h after the fall revealed femoral neck fracture of the right hip.

The patient was taken to the operating room 1 day after injury for THA of the right hip. Anteroposterior radiography (Fig. 1) and computed tomography (CT) (Fig. 2) 1 day after operation showed that periprosthetic fracture of the proximal femur involved the greater trochanter, with lateral cortical extension.

Fig. 1
figure 1

Anteroposterior radiograph of the right hip in postoperative day 1 showed the distal extension of the fracture line down the lateral cortex

Fig. 2
figure 2

a-b Computed tomography scan and three-dimensional reconstruction of the right hip in postoperative day 1 showed the distal extension of the fracture line down the lateral cortex; this leads to destabilization of the stem because the lateral buttress is lost

ORIF was performed 2 days after the diagnosis, and the 2 cables were annularly fixed above and below the small trochanter separately. Anteroposterior radiography (Fig. 3) 2 years after surgery showed the fracture healed well, and the stem was stable. Both lower limbs were of equal length, and the Harris score of the right hip was 96.

Fig. 3
figure 3

Anteroposterior radiograph 2 years after ORIF showed reduction and fixation of the fracture, the fracture healed well, and the stem is stabilized

Case 2

An 82-year-old woman suffered from severe pain in her left hip after she fell over while walking. Bilateral radiography of hips 1 h after the fall exhibited femoral neck fracture of the left hip. The patient received THA of the left hip 2 days after the injury. CT scan (Fig. 4 a) and three-dimensional reconstruction (Fig. 4 b) 1 day after the operation showed that periprosthetic fracture of the proximal femur affected the greater trochanter and the lateral cortex of the proximal femur.

Fig. 4
figure 4

a-b Computed tomography scan and three-dimensional reconstruction of the left hip in postoperative day 1 showed the distal extension of the fracture line down the lateral cortex; this leads to destabilization of the stem because the lateral buttress is lost

The patient underwent ORIF 2 days after diagnosis. Two cables were used to fix the fracture, one of which was circumferentially placed above the small trochanter and the other was placed in a ‘8-shape’ from the large trochanter to the lower trochanter. Anteroposterior radiography (Fig. 5) 2 years after ORIF showed the fracture healed well, and the stem was stable. Both lower limbs were found to be of equal length, and the Harris score of the left hip was 94.

Fig. 5
figure 5

Anteroposterior radiograph 2 years after ORIF showed reduction and fixation of the fracture, the fracture healed well, and the stem was stabilized

Discussion

The Vancouver Classification System (VCS) [6] and the Unified Classification System (UCS) [7] for PPFF have been generally accepted. The VCS focuses on the location of the fracture relative to the stem, the stability of the implant, and the associated bone loss [6]. Type A fractures are in the trochanteric region, type B fractures involve the area of the stem, and type C fractures are distant from the tip of the stem. Duncan and Haddad [7] introduced the UCS to expand and update the VCS and apply treatment principles to all periprosthetic fractures. When applied to the femur, the UCS retains the previous VCS patterns and extends to include two new fracture patterns, type D and E. Type D refers to a fracture of the femur after hip and knee arthroplasty (Type C for each joint). Type E is a fracture involving both the acetabulum and femur after hip arthroplasty.

In both VCS and UCS, Type A fractures are subdivided into fractures of the greater trochanter (AGT) and those of the lesser trochanter (ALT). Van Houwelingen and Duncan [8], and Capello et al. [9] reported pseudo ALT periprosthetic fractures that were actually Type B2 of VCS. This type of periprosthetic fracture of the lesser trochanter included a segment of the proximal medial femoral cortex. However, in this report, we presented 2 cases of periprosthetic fracture of the proximal femur involving the greater trochanter with lateral cortical extension, leading to destabilization of the stem. These periprosthetic fractures of the proximal femur involving the lesser/greater trochanter with medial/lateral cortical extension can be classified as variant Type A fractures that are actually Type B2.

On the basis of a systematic literature review and an evaluation of 402 cases of PPFF, Huang et al. [10] introduced a more precise fracture classification based on the original UCS by (1) adding two new B2 subtypes: B2PALT (i.e., pseudo ALT) and B2PAGT (i.e., pseudo AGT) and (2) adding a new FS category to encompass stem fracture, alone or accompanied with PPFF. B2PALT/B2PAGT was defined as fracture in trochanter region that includes a segment of the proximal medial/lateral femoral cortex (Fig. 6). According to the modified UCS [10], the 2 cases in this report were categorized as B2PAGT. 

Fig. 6
figure 6

The Modified Unified Classification System [10]: B2PALT/B2PAGT was defined as fracture in trochanter region including a segment of the proximal medial/lateral femoral cortex

It’s worth mentioning why this type is classified as “B2”. The key distinguishing feature between the type AGT fracture and pseudo AGT periprosthetic fracture of the greater trochanter lies in the distal extension of the fracture involving the lateral cortex of the proximal femur, which destabilizes the stem in a B2 fracture. CT scan can help clinicians to determine the stability of the stem and distinguish between fracture type A and type B. The region involved in this type of fracture (i.e., Baba classification Type 1A) can render the stem unstable [11].

The 2 variant Type AGT fractures were both diagnosed 1 day after operation. This is usually seen within 6 weeks of the index procedure, typically following the insertion of a tapered, cementless stem within a demineralized femur. The mechanism may be due to an unrecognized intraoperative fracture that is subsequently displaced under load of muscular tension, or may occur immediately after or during rehabilitation.

The principles of treatment depend on the timing of the fracture and the size of the medial/lateral fracture fragment. If recognized intraoperatively as non-propagating cortical crack, then extraction of the broach or stem, followed by cerclage cable fixation and reinsertion of the stem is adequate in most cases, plus protected weight bearing for 6 weeks. Missed diagnosis or fractures that occur in the early postoperative period with associated fracture displacement and implant subsidence often require THA revision with a longer stem, along with ORIF of the fracture using cerclage cables and/or proximal femoral plating [8, 10].

However, we did not perform a revision with a longer stem, but just employed ORIF with 2 cables, with weight bearing starting from the day after surgery. The reason why we chose ORIF alone over revision THA with longer stem lies in that (1) The mechanism of injury in variant Type AGT fractures is similar to Pseudo ALT fracture [8]. (2) Not all the Type B2 fractures require THA revision. Capello et al. [9] reported 9 Pseudo ALT fractures, whereas 3 of 9 cases were successfully managed non-surgically. In their study, the fracture had been noted early postoperatively, frequently with stem subsidence but needed no surgery and the stem restabilization, and subsequent surgery. (3) Our past experience with arthroplasty for unstable intertrochanteric osteoporotic fractures, along with ORIF of the fractures prompted us to use cerclage cables [12].

On the basis of about findings, we are led to conclude that early cerclage cable fixation alone, can successfully address this particular Vancouver Type A periprosthetic fracture variant despite reported destabilization of the femoral stem. Re-stabilization was based on the principle of stem subsidence. Although the principles of treatment suggest use of longer stem revision and the fracture fixation, ORIF has the advantages of minimal invasion and rapid recovery. In addition, we measured the stem position from the X-Ray films immediately after operation and in a two-year follow-up, and found that there was no significant stem subsidence or lower limb shortening.

Conclusion

It is important to distinguish the variant type AGT periprosthetic fracture from the type AGT, because type AGT periprosthetic fracture is associated with destabilization of the stem and requires early re-intervention. CT scan works better than X-ray examination in finding prosthetic looseness in this type of fracture. These cases illustrated that ORIF with cable could, in some variant type AGT periprosthetic fractures, achieve successful healing and stem stabilization.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Change history

  • 05 May 2020

    An amendment to this paper has been published and can be accessed via the original article.

Abbreviations

THA:

Total hip arthroplasty

ORIF:

Open reduction and internal fixation

PPFF:

Periprosthetic femoral fractures

CT:

Computed tomography

VCS:

Vancouver Classification System

UCS:

Unified Classification System (UCS)

References

  1. Malchau H, Herberts P, Eisler T, Garellick G, Söderman P. The Swedish total hip replacement register. J Bone Joint Surg Am. 2002;84-A(Suppl 2):2–20.

    Article  Google Scholar 

  2. Abdel MP, Watts CD, Houdek MT, Lewallen DG, Berry DJ. Epidemiology of periprosthetic fracture of the femur in 32 644 primary total hip arthroplasties: a 40-year experience. Bone Joint J. 2016 Apr;98-B(4):461–7.

    Article  CAS  Google Scholar 

  3. Naqvi GA, Baig SA, Awan N. Interobserver and intraobserver reliability and validity of the Vancouver classification system of periprosthetic femoral fractures after hip arthroplasty. J Arthroplast. 2012;27(6):1047–50.

    Article  Google Scholar 

  4. Lindahl H, Malchau H, Herberts P, Garellick G. Periprosthetic femoral fractures classification and demographics of 1049 periprosthetic femoral fractures from the Swedish national hip Arthroplasty register. J Arthroplast. 2005;20(7):857–65.

    Article  Google Scholar 

  5. Brady OH, Garbuz DS, Masri BA, Duncan CP. The reliability and validity of the Vancouver classification of femoral fractures after hip replacement. J Arthroplast. 2000;15(1):59–62.

    Article  CAS  Google Scholar 

  6. Duncan CP, Masri BA. Fractures of the femur after hip replacement. Instr Course Lect. 1995;44:293–304.

    CAS  PubMed  Google Scholar 

  7. Duncan CP, Haddad FS. The unified classification system (UCS): improving our understanding of periprosthetic fractures. Bone Joint J. 2014;96-B(6):713–6.

    Article  CAS  Google Scholar 

  8. Van Houwelingen AP, Duncan CP. The pseudo a (LT) periprosthetic fracture: it’s really a B2. Orthopedics. 2011;34(9):e479–81.

    PubMed  Google Scholar 

  9. Capello WN, D'Antonio JA, Naughton M. Periprosthetic fractures around a cementless hydroxyapatite-coated implant. Clin Orthop Relat Res. 2014;472(2):604–10.

    Article  Google Scholar 

  10. Huang JF, Jiang XJ, Shen JJ, Zhong Y, Tong PJ, Fan XH. Modification of the unified classification system for periprosthetic femoral fractures after hip arthroplasty. J Orthop Sci. 2018 Nov;23(6):982–6.

    Article  Google Scholar 

  11. Baba T, Homma Y, Momomura R, Kobayashi H, Matsumoto M, Futamura K, Mogami A, Kanda A, Morohashi I, Kaneko K. New classification focusing on implant designs useful for setting therapeutic strategy for periprosthetic femoral fractures. Int Orthop. 2015 Jan;39(1):1–5.

    Article  Google Scholar 

  12. Chu X, Liu F, Huang J, Chen L, Li J, Tong P. Good short-term outcome of arthroplasty with Wagner SL implants for unstable intertrochanteric osteoporotic fractures. J Arthroplasty. 2014;29(3):605–8.

    Article  Google Scholar 

Download references

Acknowledgments

None.

Funding

This work was supported by Zhejiang Provincial Natural Science Foundation (LY20H270012) and Zhejiang Administration of Traditional Chinese Medicine (2020ZB090).

Author information

Authors and Affiliations

Authors

Contributions

Study design: Huang JF and Huang Y. Study implementation: Fan MQ and Huang JF. Data collection: Chen XL and Huang JF. Drafting of the manuscript: Fan MQ and Chen XL. Approval of final version of the manuscript: Huang JF. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Yong Huang or Jie-Feng Huang.

Ethics declarations

Ethics approval and consent to participate

All authors have confirmed that this work complies with the International Committee of Medical Journal Editors (ICMJE) and the Declaration of Helsinki.

Consent for publication

Written informed consents for this publication were previously obtained from the patients.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fan, MQ., Chen, XL., Huang, Y. et al. Open reduction and internal fixation with cables for the variant AGT Periprosthetic fracture: a case report and literature review. Arthroplasty 2, 10 (2020). https://doi.org/10.1186/s42836-020-00029-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42836-020-00029-5

Keywords