Skip to main content

Horror of three synergistic factors in THA: high mechanical stress, dissimilar metals, low elasticity stem: a case report

Abstract

Background

A large-diameter femoral head is effective in preventing dislocation after total hip arthroplasty. However, although rare, catastrophic stem tribocorrosion may occur at the head-stem junction.

Case presentation

A 70-year-old woman underwent revision surgery 7.5 years after total hip arthroplasty because of catastrophic stem corrosion with dissociation of the metal head (cobalt/chromium) and stem (TiMo12Zr6Fe2). Abnormal levels of cobalt were found in the intra-articular fluid, capsule, hip muscle, and blood. Revision surgery was performed via the direct anterior approach. The well-fixed femoral stem was explanted, and a cemented stainless stem with stainless head was implanted. Three months after the revision surgery, the cobalt concentration in the blood had decreased to normal.

Conclusions

Stem dissociation in the present case might have been caused by synergistic combination of a 36-mm-diameter femoral head and long neck length offset with high frictional torque, a cobalt-chromium head with a high risk of galvanic corrosion, and a TMZF (TiMo12Zr6Fe2) alloy stem with a low Young’s modulus of elasticity. The combination of these factors must be avoided.

Background

Although total hip arthroplasty (THA) has excellent patient satisfaction and long-term survivorship, post-THA dislocation is one of its most serious complications [1]. Various attempts have been made to reduce the dislocation rate, including the surgical approach and determining the optimal implant positioning angle [2]. A large-diameter femoral head is another effective option because it increases the oscillation angle and jumping distance, thereby preventing postoperative dislocation [3, 4]. A randomized, controlled study revealed that THA with a 36-mm (instead of 28-mm) femoral head decreased postoperative dislocation [5]; however, despite its anti-dislocation effect, catastrophic stem tribocorrosion occurred at the head-stem junction (i.e., trunnionosis). The reported patient risk factors for trunnionosis are obesity, use of large metal heads, and long neck length [6].

We report a Japanese woman with abnormal cobalt levels in the intra-articular fluid, capsule, hip muscle, and blood who had early implant failure and required revision surgery 7.5 years after THA with a 36-mm metal head and TMZF (TiMo12Zr6Fe2) alloy stem. The patient provided consent for the publication of her case data.

Case presentation

A 70-year-old woman (height 165 cm, weight 75 kg) with acute right hip pain was referred to us. She had undergone THA 7.5 years previously for hip osteoarthritis. The THA had been performed via the posterior approach using a 36-mm cobalt/chromium head and 5-mm neck (Stryker Inc., Tokyo, Japan), cementless stem (Accolade TMZF Hip Stem, 127°, #3; Stryker), cementless acetabular component (50 mm Trident Acetabular shell; Stryker), and X3 polyethylene insert. Up to 6 months prior to this admission, she could walk comfortably without a cane. She and others then started to hear abnormal squeaking sounds while she was walking.

Laboratory data showed a normal C-reactive protein level (0.12 mg/dl) but abnormal serum levels of cobalt (7.0 μg/L, normal < 1 μg /L) and chromium (3.7 μg/L, normal < 5 μg /L) [7]. Radiography performed just after the previous THA and prior to this admission showed no abnormalities in cup inclination (33°) or anteversion angle (19°) (Fig. 1). However, each offset in the right hip was longer than that in the left hip (femoral/acetabular/total offset: 43.7/38.6/82.3 (right) and 42.1/33.1/75.2 (left)). Leg length was 2 mm longer on the operative side than on the healthy side [8]. Plain radiography at the time of the present admission showed implant dissociation between the stem and head, with stem tip malformation (Fig. 2). Plain computed tomography and magnetic resonance imaging showed a mass (72 × 22 × 32 mm) that was partially calcified around the hip joint. We therefore suspected femoral head-stem trunnion dissociation secondary to corrosion after THA with a large-diameter metal head that required revision surgery.

Fig. 1
figure 1

Anteroposterior plain radiography of both hips after total hip arthroplasty (THA)

Fig. 2
figure 2

Radiography obtained 7.5 years after total hip arthroplasty (THA). There is dissociation of the femoral head from the head-stem trunnion (right panel, box) after THA (right panel, black arrows), radiolucency (left panel: black arrowhead, and white arrowheads), and cortical hypertrophy (white arrow)

Surgery was performed using a direct anterior approach on a traction table [9]. Before opening the capsule, the rectus femoris and tensor fascia lata were biopsied for metal ion testing. Surgical findings included a joint capsule filled with black fluid and black granulation tissue (Fig. 3A, B), as well as thickened fibrous tissue. The anterior capsule was sent for pathological examination. Tissues debrided around the capsule were severely affected by metal ions. The pigmented granulation tissue in the joint was removed.

Fig. 3
figure 3

Intraoperative photographs. A Pigmented metal around the joint (arrow). B Black granulation tissue adhering to the inner surface of the joint capsule (arrow)

There was no loosening of the cup or stem. The removed head-stem junction was severely deformed and worn (Fig. 4). The acetabular cup was well fixed. The polyethylene liner was replaced with a metal liner for the dual mobility cup. The well-fixed femoral stem was explanted without fracture, and a cemented stem was implanted [Exeter V40 (offset 37.5 mm, # 2); Stryker] (Fig. 5). Total right hip femoral offset was intentionally decreased from 82.3 to 63.8 mm (lower than that of the left hip). Leg-length discrepancy was corrected to within 1 mm.

Fig. 4
figure 4

Photographs of the extracted instrumentation. Left panel: metal debris inside the metal head. Middle and right panels: significant deformation at the head-stem trunnion (arrows)

Fig. 5
figure 5

Postoperative radiography showing the cemented femoral stem

The microbial cultures of the joint fluid were negative. Pathological examination revealed necrotic tissue with fibrosis. Cobalt/chromium metal levels were elevated in synovial fluid, joint capsule, rectus femoris, and fascia lata (Table 1). Frozen tissue sections (5-μm thick) of the anterior joint capsule were processed. The distribution of cobalt or chromium in the joint capsule—determined by laser ablation (NWR213; Elemental Scientific Lasers, Tokyo, Japan) and inductively coupled mass spectrometry (LA-ICP-MS) (Agilent 8800; Agilent Scientific, Santa Clara, CA, USA)—is shown in Fig. 6. Concentrations were estimated using the results of conventional LA-ICP-MS after acid digestion of the neighboring tissue. Both cobalt and chromium were distributed on the inner surface of the joint capsule as small particles. Semiquantitative LA-ICP-MS images showed cobalt/chromium on the inner surface of the joint capsule (Fig. 6D).

Table 1 Distribution of cobalt and chromium ions in each living tissue biopsied in the reported patient after implant failure
Fig. 6
figure 6

Photographs of extracted specimens and distribution of cobalt or chromium in the joint capsule. A Anterior capsule. White dotted line shows the slice levels in panels B–D. B Slice of the anterior capsule. C Serial section from tissue in B (hematoxylin-eosin, × 40). D Semi-quantitative laser ablation-inductively coupled plasma mass spectrometry images. Left panels: quantitative images of cobalt or chromium. Middle panels: images of copper distribution inside the joint capsule as a position reference. Right panels: merged images of cobalt or chromium and copper distributions

Immediate alleviation of the patient’s pain was observed postoperatively, and painless walking was possible 1 month later. Her blood cobalt concentrations decreased with time and were undetectable at 3 months postoperatively (Fig. 7).

Fig. 7
figure 7

Changes in postoperative blood metal (cobalt and chromium) concentrations. M, month(s); Preop., preoperatively; op., operation

Discussion

We experienced a patient who had undergone THA with a 36-mm metal head and developed catastrophic trunnionosis with elevated blood cobalt levels that required revision surgery 7.5 years later. We concluded that the use of a 36-mm cobalt/chromium head with a dissimilar metal stem should be severely restricted.

Although previous reports suggested other risk factors—e.g., body weight, high neck offset, and activity level [10,11,12,13,14,15,16,17,18]—we believe that the three major potentially linked factors that contributed to the onset of trunnionosis in our patient were increased mechanical stress at the trunnion, contact of dissimilar metals at the trunnion, and use of an Accolade TMZF stem.

First, we think that the increased mechanical stress at the trunnion (i.e., fretting corrosion) played a major role in the development of trunnionosis. Increased mechanical stress is induced by multiple factors (e.g., greater body weight, increased activity level, longer and higher neck offset, large head size), and it is speculated that these factors synergistically generate high mechanical stress at the trunnion [10,11,12, 19]. In our case, the patient’s body weight, high neck length, longer femoral offset compared with the left hip, and large femoral head synergistically increased the mechanical stress.

Although it is difficult to determine a body mass index (BMI) cutoff value, it is obvious that heavier patients have a greater risk of increased mechanical stress at the trunnion. Previous studies have reported that the risk of trunnionosis is high in patients with a BMI > 30 kg/m2 [6], particularly men with high BMI (91.4% of trunnionosis cases occur in men) [11]. Because our patient was a woman weighing 75 kg with a BMI of 27.5 kg/m2 and no increased activity level (although the BMI was only just acceptable), the trunnionosis was likely due to other factors leading to unacceptable mechanical stress.

Although it is unclear which femoral head diameter is safest to use to avoid trunnionosis, a diameter of ≥36 mm seems to put patients at higher risk. However, trunnionosis may occur with a 32-mm femoral head if other risk factors for increased mechanical stress are present (e.g., heavy body weight, large neck offset). Dan et al reported a case of trunnionosis with a 32-mm femoral head in a patient weighing 90 kg [14]. Considering these points, surgeons should avoid unnecessary leg lengthening, an inappropriate femoral offset, and the use of a large femoral head.

Second, contact between dissimilar metals can cause galvanic corrosion due to an electrochemical process resulting from electron flow [15, 20]. Although titanium stems and cobalt/chromium heads have individual advantages, the combination of these metals could potentially cause a synergistic effect causing corrosion, especially under great mechanical stress. The implant should be selected after considering the risks and benefits of these variables.

Third, the nature of the Accolade TMZF negatively affects the onset of trunnionosis. Stems made of TMZF alloy have a low Young’s modulus of elasticity. The Young’s modulus is proportional to the flexural rigidity and that of Ti-6Al-4 V is 119 GPa while that of TMZF is 82 GPa [16]. The low flexural rigidity of TMZF is a risk factor for taper corrosion [17]. Krishnan et al reviewed modular femoral stems and found the most failures with a TMZF stem and cobalt/chromium stem combination [18]. Thus, combined ABG2 and Rejuvenate modular stems (Stryker, Kalamazoo, MI, USA) were recalled because of high revision rates due to corrosion. In addition, a systematic review reported that the Accolade TMZF poses a serious risk of trunnionosis, with trunnionosis occurring in 34/46 cases using the Accolade TMZF [11]. Although low flexural rigidity may have the advantage of normal load transmission to native bone, appropriate strength with high safety should be guaranteed.

Elevated cobalt/chromium levels are another important issue. According to Dundon et al, our patient had enough cobalt/chromium to cause multiple systemic complications, and revision surgery would be recommended even without implant failure [12]. Sometimes reported as “cobaltism,” high levels of blood cobalt cause systemic complications, including vision loss, hearing loss, cognitive decline, hypothyroidism, lymphopenia, and fatal cardiomyopathy [21,22,23,24,25]. Our patient had lower cobalt/chromium levels than reported by Ikeda et al [25], possibly because (1) the dissociation occurred before the metal concentration increased, implying the importance of early detection of trunnionosis, and/or (2) although the trunnion was massively deformed, there was minimal deformation of the metal femoral head.

Judging from the anterior capsule images of the metal concentration in the black intra-articular fluid and anterior capsule, the metal first accumulates at the joint, with the capsule a possible barrier to expansion around the hip joint. Although the cobalt blood level decreased to the normal range within 3 months after the revision surgery, metal ions were detected in the hip muscle at the time of the surgery, where there was no obvious contamination. As cobalt is thought to accumulate in strained muscle, careful and continued observation of this patient is necessary.

Conclusion

The present patient had abnormal levels of cobalt/chromium in intra-articular fluid, capsule, hip muscle, and blood, and underwent revision surgery 7.5 years after the initial THA due to catastrophic stem corrosion at the trunnion. Use of a 36-mm metal femoral head combined with a low elasticity stem must be severely restricted. Overall, the present case suggests that the systematic effects of metal should be investigated.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

THA:

Total hip arthroplasty

BMI:

Body mass index

References

  1. Charney M, Paxton EW, Stradiotto R, Lee JJ, Hinman AD, Sheth DS, et al. A comparison of risk of dislocation and cause-specific revision between direct anterior and posterior approach following elective Cementless Total hip Arthroplasty. J Arthroplast. 2020;35(6):1651–7.

    Article  Google Scholar 

  2. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60(2):217–20.

    Article  CAS  Google Scholar 

  3. Ohmori T, Kabata T, Kajino Y, Inoue D, Taga T, Yamamoto T, et al. Effect of changing femoral head diameter on bony and prosthetic jumping angles. Eur J Orthop Surg Traumatol. 2019;29(3):625–32.

    Article  Google Scholar 

  4. Sariali E, Lazennec JY, Khiami F, Catonné Y. Mathematical evaluation of jumping distance in total hip arthroplasty: influence of abduction angle, femoral head offset, and head diameter. Acta Orthop. 2009;80(3):277–82.

    Article  Google Scholar 

  5. Howie DW, Holubowycz OT, Middleton R. Large femoral heads decrease the incidence of dislocation after total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2012;94(12):1095–102.

    Article  Google Scholar 

  6. Matsen Ko L, Chen AF, Deirmengian GK, Hozack WJ, Sharkey PF. Catastrophic femoral head-stem Trunnion dissociation secondary to corrosion. J Bone Joint Surg Am. 2016;98(16):1400–4.

    Article  Google Scholar 

  7. Runner RP, Bellamy JL, Roberson JR. Gross Trunnion failure of a cobalt-chromium femoral head on a titanium stem at midterm follow-up: a report of 3 cases. JBJS Case Connect. 2016;6(4):e96.

    Article  Google Scholar 

  8. Woolson ST, Hartford JM, Sawyer A. Results of a method of leg-length equalization for patients undergoing primary total hip replacement. J Arthroplast. 1999;14(2):159–64.

    Article  CAS  Google Scholar 

  9. Homma Y, Baba T, Ozaki Y, Watari T, Kobayashi H, Ochi H, et al. In total hip arthroplasty via the direct anterior approach, a dual-mobility cup prevents dislocation as effectively in hip fracture as in osteoarthritis. Int Orthop. 2017;41(3):491–7.

    Article  Google Scholar 

  10. Norman TL, Denen JE, Land AJ, Kienitz DM, Fehring TA. Taper-Trunnion Interface stress varies significantly with head size and activity. J Arthroplast. 2019;34(1):157–62.

    Article  Google Scholar 

  11. Bansal T, Aggarwal S, Dhillon MS, Patel S. Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature. Int Orthop. 2020;44(4):609–21.

    Article  Google Scholar 

  12. Dundon JM, Ramkumar PN, Lajam C. Tribocorrosion in Total hip Arthroplasty. JBJS Rev. 2017;5(3):e5. https://doi.org/10.2106/JBJS.RVW.16.00055.

  13. Dyrkacz RM, Brandt JM, Ojo OA, Turgeon TR, Wyss UP. The influence of head size on corrosion and fretting behaviour at the head-neck interface of artificial hip joints. J Arthroplast. 2013;28(6):1036–40.

    Article  Google Scholar 

  14. Arvinte D, Kiran M, Sood M. Femoral head-Trunnion dissociation secondary to corrosion of a modular Total hip replacement: a case report. JBJS Case Connect. 2015;5(2):e35–e4.

    Article  Google Scholar 

  15. Urish KL, Giori NJ, Lemons JE, Mihalko WM, Hallab N. Trunnion corrosion in Total hip Arthroplasty-basic concepts. Orthop Clin North Am. 2019;50(3):281–8.

    Article  Google Scholar 

  16. Yang X, Hutchinson CR. Corrosion-wear of β-Ti alloy TMZF (Ti-12Mo-6Zr-2Fe) in simulated body fluid. Acta Biomater. 2016;42:429–39.

    Article  CAS  Google Scholar 

  17. Kao YY, Koch CN, Wright TM, Padgett DE. Flexural rigidity, taper angle, and contact length affect fretting of the femoral stem Trunnion in Total hip Arthroplasty. J Arthroplast. 2016;31(9 Suppl):254–8.

    Article  Google Scholar 

  18. Krishnan H, Krishnan SP, Blunn G, Skinner JA, Hart AJ. Modular neck femoral stems. Bone Joint J. 2013;95-b(8):1011–21.

    Article  CAS  Google Scholar 

  19. Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–61.

    Article  Google Scholar 

  20. Collier JP, Surprenant VA, Jensen RE, Mayor MB, Surprenant HP. Corrosion between the components of modular femoral hip prostheses. J Bone Joint Surg Brit. 1992;74(4):511–7.

    Article  CAS  Google Scholar 

  21. Gessner BD, Steck T, Woelber E, Tower SS. A systematic review of systemic Cobaltism after Wear or corrosion of chrome-cobalt hip implants. J Patient Saf. 2019;15(2):97–104.

    Article  Google Scholar 

  22. Garcia MD, Hur M, Chen JJ, Bhatti MT. Cobalt toxic optic neuropathy and retinopathy: case report and review of the literature. Am J Ophthalmol Case Rep. 2020;17:100606.

    Article  Google Scholar 

  23. Umar M, Jahangir N, Faisal Khan M, Saeed Z, Sultan F, Sultan A. Cobalt cardiomyopathy in hip arthroplasty. Arthroplast Today. 2019;5(3):371–5.

    Article  Google Scholar 

  24. Hart AJ, Skinner JA, Winship P, Faria N, Kulinskaya E, Webster D, et al. Circulating levels of cobalt and chromium from metal-on-metal hip replacement are associated with CD8+ T-cell lymphopenia. J Bone Joint Surg Brit. 2009;91(6):835–42.

    Article  CAS  Google Scholar 

  25. Ikeda T, Takahashi K, Kabata T, Sakagoshi D, Tomita K, Yamada M. Polyneuropathy caused by cobalt-chromium metallosis after total hip replacement. Muscle Nerve. 2010;42(1):140–3.

    Article  Google Scholar 

Download references

Acknowledgments

We thank Kelly Zammit, BVSc, from Edanz (https://jp.edanz.com/ac) for editing a draft of this manuscript.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

Seiya Ishii: wrote the initial draft. Yasuhiro Homma: wrote and revised the paper. Takehisa Matsukawa: performed the data analysis and interpretation for metal concentration, ensured the accuracy of the data, revised the paper. Tomonori Baba, Kazuo Kaneko, Muneaki Ishijima: revised the paper. Ayano Kubota: performed the frozen tissue sectioning, and the data analysis and interpretation for metal concentration. Kazuhito Yokoyama: performed the data analysis and interpretation for metal concentration, ensured the accuracy of the data, revised the paper. The authors read and approved the final manuscript.

Corresponding author

Correspondence to Yasuhiro Homma.

Ethics declarations

Ethics approval and consent to participate

All authors have confirmed that this work complies with the International Committee of Medical Journal Editors (ICMJE) and the Declaration of Helsinki.

Consent for publication

Written informed consent for the publication of this case report was obtained from the patient.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ishii, S., Homma, Y., Matsukawa, T. et al. Horror of three synergistic factors in THA: high mechanical stress, dissimilar metals, low elasticity stem: a case report. Arthroplasty 3, 39 (2021). https://doi.org/10.1186/s42836-021-00091-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42836-021-00091-7

Keywords