Does robotic technology successfully restore the joint line after total knee arthroplasty? A retrospective analysis

Abstract

Background

Our study aims to determine the effectiveness of robotic technology for total knee arthroplasty in the successful restoration of the joint line of the knee with respect to that of a normal human anatomical knee. The restoration of the joint line is an important technical goal on which the postoperative outcomes and the success of the surgery depend.

Methods

Sixty-four postoperative plain anteroposterior radiographs of 60 patients, who received total knee arthroplasty by using the robotic technology were analyzed and compared with 66 similar radiographs of 60 patients who received the conventional method. The distances of the lateral epicondyle to the joint line (LEJL) and proximal tibiofibular joint to the joint line (PTFJJL) were calculated and analyzed.

Results

We found that the mean value of LEJL minus PTFJJL in the robotic group was 0.334 ± 0.115 (mean ± SD), while in the conventional group, it was 2.304 ± 0.308. The difference between the two groups was statistically significant. The mean ratio (LEJL:PTFJJL) in the robotic group was also equal to 1.017 ± 0.042.

Conclusion

From these findings it could be concluded that the robotic technology significantly increases the accuracy of the total knee arthroplasty and, compared to the conventional method, achieves an almost anatomical position of the joint line.

Background

Restoring joint line is an important goal of total knee arthroplasty (TKA) [1]. Failure to restore the joint line to its anatomical position can result in anterior knee pain, mid-flexion instability, reduced range of motion, and patellar mal-tracking [1]. All of this leads to lower knee scores and lower patient satisfaction. The success of a total knee arthroplasty depends upon the restoration of the normal knee kinematics. Various modifications to conventional methods, implant designs, and instruments are in an ongoing process to achieve the same goals. Many computer-assisted and navigational devices have been introduced to date. One of such advances is the introduction of robotic technology. This technology serves as an intraoperative guidance tool for the operating surgeon and allows foran objective assessment at every step of the procedure, thus, increasing the precision of the surgery [2]. Not many studies have been carried out to analyze these robotic systems with regard to their effectiveness and their contribution to the success of the surgery. The aim of this retrospective study was to analyze how precisely the joint line is restored using robotic technology for the primary total knee replacement.

Methods

Plain anteroposterior (AP) radiographs of 64 knees from 60 patients who had undergone total knee replacement using robotic technology between November 2017 and March 2020 were analyzed and compared with the AP radiographs of 66 knees of 60 patients who had undergone a TKA using the conventional method during the same period. The consent of all the patients was obtained for the study.

Inclusion criteria: (1) Cases who received total knee arthroplasty by using robotic technology  or conventional methods; (2) Those who were preoperatively diagnosed as having primary knee osteoarthritis with genu varus deformity and fixed flexion deformity of less than 15°; (3) Patients whose true postoperative AP radiographs were available.

Exclusion criteria: (1) Subjects who were preoperatively diagnosed with post-traumatic osteoarthritis and inflammatory arthritis; (2) Those who had preoperative valgus knee deformity; (3) Candidates who had pre-existing hip pathology and hip arthroplasty.

The operative procedure, using robotic technology, begins with connecting the hardware, selecting bur and control, and calibrating the handpiece. Surgical choice was according to surgeon's  preferences. We used the anterior referencing system, the posterior condylar axis as the reference axis for rotation and the resection depth equal to the implant thickness. The exposure was achieved by using the medial parapatellar approach and patelloplasty was done. Trackers were attached and bony landmarks were marked for the femur and tibia. The hip center was calculated. The preoperative range of motion of the knee was recorded. Afterwards, the femoral and tibial articular surfaces were then mapped. Virtual images were then used to plan on the screen. We aimed at achieving < 1–2 mm of joint laxity in the lateral compartment over the entire range of motion of the knee. Femoral and tibial cuts were then made with a 5 mm bur according to the final plan. Trials were then taken and the sizes were confirmed. Cementation and final implantation were  performed. The robotic system (Smith & Nephew Navio PFS, Blue Belt Technologies, Plymouth, MN)is a haptic system. The implants used were Anthem or Genesis II, CR (cruciate retaining)/PS (posterior-cruciate sacrificing) knee implants.

The proximal tibiofibular joint (PTFJ) and the lateral epicondyle (LE) of the femur were used as anatomical reference points for the analysis of the joint lines on the X-ray images. The PTFJ can be identified as the center of the horizontal portion of the proximal tibiofibular joint [3]. If it is not visible due to fibular rotation, it is at the intersection of the lateral prominence of the fibular head and the fibular styloid [3, 4]. LE is the most prominent bony point on the lateral distal femur, from which the lateral collateral ligament originates [3, 4]. The joint line (JL) is represented in the AP X-ray image by a line drawn tangentially to the distal surface of the femoral prosthesis. The distances of the joint line from the lateral epicondyle of the femur (LEJL) and the proximal tibiofibular joint (PTFJJL) were measured and compared. If the joint line is at the same distance from these two bony landmarks, it can be said that the joint line was successfully restored [4]. All data were calculated and collected by the corresponding author only. SCANDOC DICOM 5.2.0.0 and MICRODICOM 3.1.4 software packages were used as calculation software.

We analyzed the postoperative functional outcome of the surgery in both the robotic and conventional groups using the Oxford Knee Score (OKS), which consists of 12 questions [5,6,7]. The score rates the pain status and activity level of the knee, as reported by the patient. Each question was rated between 0 (the worst outcome) and 4 (the best outcome). All scores were summed up to give the final score which ranged between 0 (the worst outcome) and 48 (the best outcome) [5,6,7].

The probability value < 0.05 was considered as a significant level for all the statistical analyses. The results were analyzed using an unpaired t-test and ANOVA. The level of evidence of our study is 3.

Results

There was no significant difference in the demographic data between the patients operated with the robotic technology (R-TKA) and the conventional method (C-TKA) (Table 1). Forty-four of 64 joints (68.75%) in the R-TKA group had absolute values of LEJL minus PTFJJL < 1 mm. The value of LEJL minus PTFJJL in this group was 0.334 ± 0.115 (mean ± SD) (Table 2). Using the unpaired t-test, this difference was found to be insignificant (P = 0.41). The value of (LEJL:PTFJJL) in the robotic group was 1.017±  0.042. Thus, with both methods, we found that both the distances are nearly equal, which suggests that the joint line lies equidistant from both these two points, which confirms the accuracy of the robotic method. The value of LEJL minus PTFJJL in the conventional group was 2.304 ± 0.308. The difference between the mean values in the robotic and conventional groups turned out to be highly statistically significant (P < 0.001) (Table 2). We also found that the precision of this method was independent of the age, gender, body mass index (BMI), and side of the operated patient (Tables 3, 4, 5, 6). We divided the data into different comparison groups within these demographic categories and analyzed the means of LEJL minus PTFJJL in these groups. We found that the difference was not significant. Figure 1a and b show the method used for calculating the LEJL and PTFJJL distances.

Table 1 Comparison of demographic characteristics between the R-TKA and C-TKA groups (N.S. = not significant)
Table 2 Comparison of the mean values of LEJL minus PTFJJL between the patients receiving robotic technology and those receiving the conventional method (H.S. = highly significant)
Table 3 Comparison between the mean values of LEJL minus PTFJJL in different age groups of the R-TKA group
Table 4 Comparison between the mean values of LEJL minus PTFJJL with respect to BMI in the R-TKA group
Table 5 Comparison of mean values of LEJL minus PTFJJL between male and female genders in the R-TKA group
Table 6 Comparison of mean values of LEJL minus PTFJJL with respect to right or left knee operated in the R-TKA group
Fig. 1
figure 1

a Plain AP radiograph of a patient’s left knee from the sample showing how the LEJL and PTFJJL distances were calculated. The horizontal line is the joint line achieved after the surgery. The perpendicular above this line is LEJL and the perpendicular below this line is PTFJJL. Both these distances are nearly equalb Sketch diagram of a left knee showing how the LEJL and PTFJJL distances were calculated. The horizontal line is the joint line (red) of the knee. The perpendicular above this line is LEJL (blue) and the perpendicular below this line is PTFJJL (green). Both these distances are nearly equal

In the postoperative analysis with the OKS scoring system, we found that 12 months after operation,  in the R-TKA group, 68.25% of the knee scores were between 40 and 48 and 31.75% of the scores were between 30 and 39. While in the C-TKA group, 51.61% of the knee scores were between 40 and 48 and 48.39% of the scores were between 30 and 39. The mean OKS value in the R-TKA group was found to be 40.87 ± 4.675, while in the C-TKA group, the mean OKS value was found to be 37.65 ± 5.125. We lost 3 cases to follow-up in the R-TKA group and 4 cases in the C-TKA group.

Discussion

The results of our study are broadly in line with what was reported by Gavin et al [3] According to their findings, the joint line of a normal knee is equidistant from the LE and PTFJ with a mean of 1.0 ± 0.1 for the ratio of LEJL:PTFJJL. Figgie et al [8] had first suggested the importance of restoring joint line in their study and found that a variation in joint line of more than 8 mm can lead to poorer results in primary total knee arthroplasty (TKA). Similarly, in their 1999 study, Partington et al [9] suggested  < 8 mm as an acceptable variation in the joint line for better results. While Hofmann et al [10] suggested in 2006 that the elevation or depression of the joint line should not be more than 4 mm. In our study, we took the joint line in a normal human anatomical knee as described by Gavin et al [3] as a reference, and found that the maximum variation in the joint line after robotic-assisted primary TKA is < 2.5 mm compared to the reference, with 68.75% of the cases showing a deviation of < 1 mm and 26.56% of the cases showing a deviation of 1–2 mm. Three cases (4.68%) showed variations between 2 and 2.5 mm.

The success of TKA depends upon soft tissue balancing, bone cuts, cementation, and restoration of the joint line. The knee joint is a hinge and pivot type of joint with 6 degrees of freedom: 3 rotations and 3 translations. Any change in the joint line and angle can lead to joint instability and failure of the surgery [1, 11]. As shown in previous studies, a joint line variation of more than 8 mm [8, 9, 12] or more than 4 mm [10] leads to poorer results. Chiu et al reported that elevation of the joint line by more than 10 mm leads to a reduction in the flexion range by more than 25% [1, 13]. Restoring the joint line to the near-normal results in improved knee scores [9]. Amarnath et al [4] mentioned that conventional methods lack accuracy and lead to the mal-positioned and frequently elevated joint line. Elevation of the joint line leads to many problems such as anterior knee pain, reduced range of motion (ROM), patella baja, mid-flexion instability [14], patellar tendon impingement, and accelerated wear [1]. Hence, it is important to restore the joint line near normal for better outcomes of the procedure.

Several methods have been described for analyzing the joint line [4]. Earlier studies attempted to analyze the joint line,  taking into account various bone and soft tissue landmarks and ratios [3, 4, 15,16,17,18]. Rajagopal et al [16] and Mountney et al [17] suggested that the ratio between inter-epicondylar distance (IED) and perpendicular to the joint line from the medial epicondyle or lateral epicondyle is constant and is about 3.0. However, Servien et al [18] found that this distance was variable depending upon the size and gender of the individual. He also found that there is a wide variation in the bony landmarks. The medial epicondyle is a sulcus between two protuberances and thus an inconsistent landmark [3, 4]. LEJL is therefore regarded as the most reliable distance for assessing the restoration of the joint line [3, 4]. There is no consensus on the exact reference point on the fibular head [19, 20]. Also, the fibular styloid has a variable morphology [3, 4, 18]. Therefore, Amarnath et al [4] and Galvin et al [3] suggested, from their studies, that the most effective method of analyzing a successful restoration of the joint line is to determine the distance of the joint line from LE and PTF joint. If these two distances are nearly equal, it confirms the successful restoration of the joint line [3, 4]. We used this method to analyze the joint line.

The robotic system enables the surgeon to assess the knee kinematics, in all 6-degrees of freedom of movement on the screen, for instance, the ROM, mediolateral laxity, and alignment of the joint over the entire ROM and also the three-dimensional view of the condyles of the femur and tibia by collecting the femoral and tibial collection points. Depending on the surgeon’s preference, he or she can customize the settings for cruciate ligament-retaining or cruciate ligament-substituting methods before the start of the procedure. The system gives guidance about the further steps of the procedure accordingly. The complete planning of the cuts to be made, the position of the implants, the size of the components, the resulting limb alignment, mediolateral laxity, and the ROM of the knee, that will be achieved at the end of the implantation, can be calculated on the planning screen before any cuts are made. Any femoral notching, component overhang, or malrotation is easily noticeable on the screen. The cutting zigs are also placed under the guidance of the system. The use of the system thus avoids the excessive soft tissue dissection required to meet the surgeon’s need for exposure and also avoids the over- or under-resection of the bone to compensate for the inadequate soft tissue releases and vice versa, as seen in the conventional method in which all these measurements are taken manually by the surgeon using the available guides from the implant system. The final results achieved after the implant has been positioned can also be analyzed on the robotic system in order to rule out errors during the implantation. This technology therefore combines both measured resection and gap balancing techniques and the surgeon does not have to rely on a single technique. In the conventional method, the analysis is dependent more on the surgeon’s subjective findings and therefore increases the likelihood of interpersonal errors. The accuracy of a surgeon’s analysis in the conventional method increases with his experience and knowledge. In the case of robotic technology, however, once a surgeon is familiar with the system, the analysis is objective, as shown on the screen, and its accuracy does not vary with the surgeon’s experience and therefore does not affect patient outcomes.

Recent studies also showed that the use of robotic technology does not lead to increased surgical time or complications [21,22,23,24,25]. Coon et al have come to the conclusion that once the surgeon gets familiar with the robotic technology, the operative time is reduced to < 40 min [26]. Robotic technology is intended to assist the operating surgeon and not to replace him [27]. The use of robots in the medical field dates back to the late 1980s [28]. Robodoc (Curexo Technology, Fremont, CA) was the first robot to be used in total hip and knee arthroplasty in 1992 [21]. However, the versatility of this system was less, which, along with increased operative time and complexity of the system, could not make it very popular amongst the surgeons [21, 29]. To overcome these shortcomings, many new robotic systems have been developed that allow for greater precision in femoral and tibial cuts. Navio PFS (Blue Belt Technologies, Plymouth, MN) is one of such robotic systems. In contrast to navigation systems, robotic systems use programmable devices that provide the surgeon with continuous feedback on the spatial position and orientation of the instruments, guiding the procedure, and following the surgical plan. This helps in achieving better kinematic alignment, mechanical alignment [24, 3031], and soft tissue balancing [2, 21, 28, 30]. This, in turn, helps to achieve better implant survival, better knee function, and better patient-reported outcomes [22].

Robotic-assisted knee replacement technology helps reduce postoperative pain [24, 30,31,32,33], restore knee function, and thus improve the quality of life in patients with severe osteoarthritis [34]. Khlopas et al [35] and Kayani et al [25] have suggested, in their studies, that these systems also reduce excessive soft tissue trauma during surgery compared to the conventional systems and thus reduce the postoperative pain. According to a retrospective study involving 3100 navigated TKA, the prevalence of fractures at the pin-insertion site is 0.16% and infection at the pin-site is 0.47% [36]. However, the data on the prevalence of these complications in robotic systems are still insufficient. To date, robotic-assisted TKA has not been shown to be cost-effective due to several factors involved [37]. However, it can become cost-effective when used in high volume centers where the major capital is recovered in the first year [22, 37, 38] and if it averts the risk of revision surgery [21, 39]. The effective benefits of this technology are still being explored and, once proven, can help determine the risk of revision surgery following robotic-assisted knee replacement surgery.

Despite the above results, our study still have a few limitations. First, in this study we could not compare the postoperative Oxford Knee Score to the preoperative score because the preoperative scoring data were not available as the study was retrospective. Secondly, there was a deviation of 0.01–0.1 degrees in the perpendiculars drawn from the LE and PTFJ to the joint line (software errors). However, this variation was very small and did not influence the overall observations.

Conclusion

We were thus able to deduce from our study that the use of robotic technology in total knee replacement surgery achieves a nearly anatomical position of the joint line and the accuracy of the surgery can be significantly increased compared to the conventional method. The effects on patient-related outcomes and the need for revision surgery compared to the conventional method need to be further analyzed.

Availability of data and materials

The data sets used and / or analyzed in the current study are available from the corresponding author on reasonable request.

Abbreviations

LEJL:

Lateral epicondyle to the joint line

PTFJJL:

Proximal tibiofibular joint to the joint line

TKA:

Total knee anthroplasty

AP:

Anteroposterior

CR:

Cruciate-retaining

PS:

Posterior-cruciate sacrificing

PTFJ:

Proximal tibiofibular joint

LE:

Lateral epicondyle

JL:

Joint line

R-TKA:

Robotic total knee anthroplasty

C-TKA:

Conventional total knee anthroplasty

BMI:

Body mass index

ROM:

Range of motion

IED:

Inter-epicondylar distance

OKS:

Oxford Knee Score

References

  1. Kumar N, Mukhopadhaya J, Yadav C, Anand S, Singh S. Joint line restoration in total knee arthroplasty. J Orthop Allied Sci. 2017;5:10–4.

    Google Scholar 

  2. Bautista M, Manrique J, Hozack WJ. Robotics in total knee arthroplasty. J Knee Surg. 2019;32(7):600–6.

    Article  Google Scholar 

  3. Pereira GC, von Kaeppler E, Alaia MJ, Montini K, Lopez MJ, Di Cesare PE, et al. Calculating the position of the joint line of the knee using anatomical landmarks. Orthopedics. 2016;39(6):381–6.

    Article  Google Scholar 

  4. Amarnath R, Thirunarayanan V, Kumar S, Kumar GV. Calculating position of joint line of knee using various radiological parameters based on Indian population. IJOS. 2017;3(3):1118–23.

    Google Scholar 

  5. Dawson J, Fitzpatrick R, Carr A, Murray D. Questionnaire on the perceptions of patients about total hip replacement. J Bone Joint Surg Br. 1996;78(2):185–90.

    Article  CAS  Google Scholar 

  6. Dawson J, Fitzpatrick R, Murray D, Carr A. Questionnaire on the perceptions of patients about total knee replacement. J Bone Joint Surg Br. 1998;80(1):63–9.

    Article  CAS  Google Scholar 

  7. Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, et al. The use of the Oxford hip and knee scores. J Bone Joint Surg Br. 2007;89(8):1010–4.

    Article  CAS  Google Scholar 

  8. Figgie HE 3rd, Goldberg VM, Heiple KG, Moller HS 3rd, Gordon NH. The influence of tibial-patellofemoral location on the function of the knee in patients with the posterior stabilized condylar knee prosthesis. J Bone Joint Surg Am. 1986;68:1035–40.

    Article  Google Scholar 

  9. Partington PF. Joint line restoration after revision total knee arthroplasty. Clin Orthop Relat Res. 1999;367:165–71.

    Article  Google Scholar 

  10. Hofmann AA, Kurtin SM, Lyons S, et al. Clinical and radiographic analysis of accurate restoration of the joint line in revision total knee arthroplasty. J Arthroplast. 2006;21(8):1154–62. https://doi.org/10.1016/j.arth.2005.10.026.

    Article  Google Scholar 

  11. Howell SM, Howell SJ, Kuznik KT, Cohen J, Hull ML. Does a kinematically aligned total knee arthroplasty restore function without failure regardless of alignment category? Clin Orthop Relat Res. 2013;471:1000–7.

    Article  Google Scholar 

  12. Laskin RS. Joint line position restoration during revision total knee replacement. Clin Orthop Relat Res. 2002:169–71.

  13. Chiu KY, Ng TP, Tang WM, Yau WP. Review article: knee flexion after total knee arthroplasty. J Orthop Surg (Hong Kong). 2002;10:194–202.

    Article  CAS  Google Scholar 

  14. Martin JW, Whiteside LA. The influence of joint line position on knee stability after condylar knee arthroplasty. Clin Orthop Relat Res. 1990:146–56.

  15. Clement, Nicholas D, David F Hamilton, Richard Burnett. A technique of predicting radiographic joint line and posterior femoral condylar offset of the knee. Arthritis. 2014.

  16. Rajagopal, T & Nathwani, Dinesh. Can interepicondylar distance predict joint line position in primary and revision knee arthroplasty? American journal of orthopedics (Belle Mead, N.J.). 2011; 40. 175–8.

  17. Mountney, John & Karamfiles, Richard & Breidahl, William & Farrugia, Mark & Sikorski, Jerzy. The Position of the Joint Line in Relation to the Trans-Epicondylar Axis of the Knee. The Journal of arthroplasty. 2008; 22. 1201–7. https://doi.org/10.1016/j.arth.2006.11.023.

  18. Servien E, Viskontas D, Giuffrè BM, Coolican MR, Parker DA. Reliability of bony landmarks for restoration of the joint line in revision knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2008;16:263–9.

    Article  Google Scholar 

  19. Espregueira-Mendes JD, da Silva MV. Anatomy of the proximal tibiofibular joint. Knee Surg Sports Traumatol Arthrosc. 2006;14(3):241–9.

    Article  CAS  Google Scholar 

  20. Havet E, Gabrion A, Leiber-Wackenheim F, Vernois J, Olory B, Mertl P. Radiological study of the knee joint line position measured from the fibular head and proximal tibial landmarks. Surg Radiol Anat. 2007;29(4):285–9.

    Article  Google Scholar 

  21. Jacofsky DJ, Allen M. Robotics in arthroplasty: a comprehensive review. J Arthroplast. 2016;31(10):2353–63.

    Article  Google Scholar 

  22. van der List JP, Chawla H, Joskowicz L, Pearle AD. Current state of computer navigation and robotics in unicompartmental and total knee arthroplasty: a systematic review with meta-analysis. Knee Surg Sports Traumatol Arthrosc. 2016;24(11):3482–95.

    Article  Google Scholar 

  23. Liow MHL, Goh GS, Wong MK, Chin PL, Tay DK, Yeo SJ. Robotic-assisted total knee arthroplasty may lead to improvement in quality-of-life measures: a 2-year follow-up of a prospective randomized trial. Knee Surg Sports Traumatol Arthrosc. 2017;25(09):2942–51.

    Article  Google Scholar 

  24. Song EK, Seon JK, Park SJ, Jung WB, Park HW, Lee GW. Simultaneous bilateral total knee arthroplasty with robotic and conventional techniques: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc. 2011;19(07):1069–76.

    Article  Google Scholar 

  25. Kayani B, Konan S, Pietrzak JRT, Haddad FS. Iatrogenic bone and soft tissue trauma in robotic-arm assisted total knee arthroplasty compared with conventional jig-based total knee arthroplasty: a prospective cohort study and validation of a new classification system. J Arthroplast. 2018;33(08):2496–501.

    Article  Google Scholar 

  26. Coon TM. Integrating robotic technology into the operating room. Am J Orthop. 2009;38(2, Suppl):7–9.

    PubMed  Google Scholar 

  27. Davies B. A review of robotics in surgery. Proc Inst Mech Eng. 2000;214:129–40.

    Article  CAS  Google Scholar 

  28. LangJE MS. FloydAJ, et al. robotic systems in orthopaedic surgery. J Bone Joint Surg Br. 2011;93(10):1296–9.

    Article  Google Scholar 

  29. Parsley BS. Robotics in orthopedics: a brave new world. J Arthroplast. 2018;33(08):2355–7.

    Article  Google Scholar 

  30. Liow MH, Xia Z, Wong MK, Tay KJ, Yeo SJ, Chin PL. Robot-assisted total knee arthroplasty accurately restores the joint line and mechanical axis. A prospective randomized study. J Arthroplast. 2014;29(12):2373–7.

    Article  Google Scholar 

  31. Song EK, Seon JK, Yim JH, Netravali NA, Bargar WL. Robotic-assisted TKA reduces postoperative alignment outliers and improves gap balance compared to conventional TKA. Clin Orthop Relat Res. 2013;471(01):118–26.

    Article  Google Scholar 

  32. Marchand R, Sodhi N, Khlopas A, et al. Patient satisfaction outcomes after robotic- arm-assisted total knee arthroplasty: a short-term evaluation. J Knee Surg. 2017;30(09):849–53.

    Article  Google Scholar 

  33. KayaniB KS. TahmassebiJ, PietrzakJRT, HaddadFS. Robotic-arm assisted total knee arthroplasty is associated with improved early functional recovery and reduced time to hospital discharge compared with conventional jig-based total knee arthroplasty. Bone Joint J. 2018;100-B(07):930–7.

    Article  Google Scholar 

  34. ShanL SB. SuzukiA, NouhF, SaxenaA. Intermediate and long-term quality of life after total knee replacement: a systematic review and meta-analysis. J Bone Joint Surg Am. 2015;97(02):156–68.

    Article  Google Scholar 

  35. KhlopasA CM, HamppEL, et al. Robotic-arm-assisted total knee arthroplasty demonstrated soft tissue protection. Surg Technol Int. 2017;30:441–6.

    Google Scholar 

  36. Brown MJ, Matthews JR, Bayers-Thering MT, Phillips MJ, Krackow KA. Low incidence of postoperative complications with navigated total knee arthroplasty. J Arthroplast. 2017;32(07):2120–6.

    Article  Google Scholar 

  37. Swank ML, Alkire M, Conditt M, Lonner JH. Technology and cost-effectiveness in knee arthroplasty: computer navigation and robotics. Am J Orthop. 2009;38(2, Suppl):32–6.

    PubMed  Google Scholar 

  38. Moschetti WE, Konopka JF, Rubash HE, Genuario JW. Can robot-assisted unicompartmental knee arthroplasty be cost-effective? A Markov decision analysis. J Arthroplast. 2016;31(04):759–65.

    Article  Google Scholar 

  39. Lonner JH, Fillingham YA. Pros and cons: a balanced view of robotics in knee arthroplasty. J Arthroplast. 2018;33(07):2007–13.

    Article  Google Scholar 

Download references

Acknowledgments

We would like to thank Dr. Nidhi Nagar (Research Project Coordinator), for helping us with the approval of the Ethics Committee and Institutional Review Board. We thank the patients who participated in our study. We would also like to thank the staff at Lokmanya Hospital (Pune City, Maharashtra State, India), who contributed to our study.

Funding

This research did not receive any specific grants from funding agencies in the public, commercial or non-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

VA: - Conceptualization, Methodology, Software, Validation, Formal analysis, Investigation, Data Curation, Writing, Visualisation, Supervision. NV: - Conceptualization, Methodology, Validation, Resources, Supervision. AG: - Methodology, Investigation, Formal analysis, Writing, Data Curation. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Varun O. Agrawal.

Ethics declarations

Ethics approval and consent to participation

Approval from the Ethics Committee and Institutional Review Board for this study was obtained from Lokmanya Medical Research Center, Pune. The reference number is LMRC / IEC / 054.

Consent to participation and publication

Written informed consent of the patients was obtained for participation and publication.

Competing interests

The authors declare that they have no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Agrawal, V.O., Gadekar, A.P. & Vaidya, N. Does robotic technology successfully restore the joint line after total knee arthroplasty? A retrospective analysis. Arthroplasty 4, 6 (2022). https://doi.org/10.1186/s42836-021-00103-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42836-021-00103-6

Keywords